The Handy Polaris hand-held meter makes it easy to measure dissolved oxygen and temperature. The flat foil pushbuttons are easy to operate and the instrument is always ready to use – just press the “on” button and measure.
Dissolved oxygen values are shown both in mg/l (ppm) and % saturation – you can choose which parameter is shown in large figures. Measurements are compensated for changes in barometric pressure.
Functions such as automatic calibration and self-check of the probe and meter combine with the well-known accuracy, reliability and low maintenance needs of the Handy series to make an instrument that will serve you well for many years to come.
Technical Advantages
Meter
- Automatic calibration with stability check.
- Automatic self-check of instrument, cable and probe.
- Compensates for changes in atmospheric pressure
- User selectable language and main display parameter.
- “On-screen” instructions.
- Waterproof.
- 1400 hours use from one 9V alkaline battery.
- Illuminated display with variable intensity.
- Automatic shutdown for power saving.
Probe
- Galvanic type.
- Self-polarizing.
- No warm up time.
- Fully temperature compensated.
- Remarkably short response time.
- Stores dry.
- Excellent long-term stability.
- Tough membrane.
- No need for regular service or renovation.
- Probe renovation easy and fast when needed.
- Spares available at very low price
Specifications
- Display: Large, easy to read graphical LCD display. Variable backlight.
- Probe Type: Galvanic cell, self-polarising, self-temperature compensating.
- Cable Length: Standard 3 m. Any length available on request.
- Operating Conditions: Probe: -5 to +45°C, meter -20 to +60°C.
- Encapsulation (meter): Short-term immersion proof to max. 5m depth.
- Measuring Range: 0-60.0 ppm (mg/l) and 0-600% saturation. Automatically compensated for temperature and barometric pressure. Manual salinity compensation.
- Temperature -5 to +45°C.
- Salinity Compensation: 0-59 ppt salinity (manually set).
- Accuracy: Depends on calibration accuracy. Typically better than +/- 1% of measured value +/- 1 digit within the standard measuring range.
- Repeatability: Typically better than +/- 0.5% of measured value.
- Response Time: 90% of end value in less than 20 seconds in water.
- Accuracy, Temperature: +/- 0.2°C.
- Dimensions: 98 mm diameter x 36 mm + belt clip (if used).
- Self-check of: Probe function, meter function, cable, battery.
- Useful battery life: Typically 500 hours normal use, max. 2 years.
- Standard Accessories: Membranes with O-rings, electrolyte, cathode cleaning pad, storage pouch